September 1, 2015
June 1, 2015
March 10, 2015
New product Vimo-151
Technical description
1. Purpose of the device
VIMO-151 is a compact, automatic hardware and software package to determine the concentration of a substance in a fluid (NaOCl Sodium hypochlorite) the indirect method, based on the density measurement. The main application - in industrial processes.
VIMO-151 consists of a membrane pressure sensors combined in a pair, to determine the difference in head pressure of the medium in two points. Sensors mounted on the vertical sampling vessel at a fixed distance. A temperature sensor is installed in the middle tank to compensate for ambient temperature variations in the process.
1. Purpose of the device
VIMO-151 is a compact, automatic hardware and software package to determine the concentration of a substance in a fluid (NaOCl Sodium hypochlorite) the indirect method, based on the density measurement. The main application - in industrial processes.
VIMO-151 consists of a membrane pressure sensors combined in a pair, to determine the difference in head pressure of the medium in two points. Sensors mounted on the vertical sampling vessel at a fixed distance. A temperature sensor is installed in the middle tank to compensate for ambient temperature variations in the process.
By sampling container is attached a flexible inlet tube for injecting the sample fluid pump and a flexible exhaust tube to drain fluid sample after the measurement. Control of the injection and discharge of fluid is performed by two pinch solenoid valves.
Sensors and actuators are connected to the computing unit. On the block is LCD touch panel for instrument control and display of measurement parameters. All elements VIMO-151 mounted on an aluminum frame, which can be installed on a horizontal and on a vertical plane.
Measurements are performed automatically discretely by selecting the liquid sample from the workflow.
1. Vertical sampling capacitance
2. Pressure Sensor
3. Temperature Sensor
4. The inlet tube
5. The discharge pipe
6. pump
7. The computing unit
8. Touch LCD control panel
9. Aluminum frame
2. Features
Power supply: 220V or 24V built-in battery
Measurement accuracy: less than 0.1% (density)
Cyclical measurement: 2 minutes
Medium: liquids, including aggressive
Membrane material: titanium
Control: LCD panel
Communication: Ethernet / Modbus TCP
Max. lift height of the intake tube 3 m
Class of protection: IP65 / IP68
3. Comparison to other products
1. Vertical sampling capacitance
2. Pressure Sensor
3. Temperature Sensor
4. The inlet tube
5. The discharge pipe
6. pump
7. The computing unit
8. Touch LCD control panel
9. Aluminum frame
2. Features
Power supply: 220V or 24V built-in battery
Measurement accuracy: less than 0.1% (density)
Cyclical measurement: 2 minutes
Medium: liquids, including aggressive
Membrane material: titanium
Control: LCD panel
Communication: Ethernet / Modbus TCP
Max. lift height of the intake tube 3 m
Class of protection: IP65 / IP68
3. Comparison to other products
Feature / Similar products
|
CLT10
(Hach, U.S.) |
Density meter DT303 (Smar, Brasil)
|
AGH-3,
Lumex ltd., RU
|
Vimo-151,
Atria ltd., RU
|
Measurement method
|
amperometric
|
capacitive indirect
|
optic
|
capacitive indirect
|
Range
|
0 – 20 mg/l
|
500 — 5000 kg/m3 *
|
20 - 300 g/l
|
20 - 500 g/l,
0-2000 kg/m3 *
|
Process environment
|
solution of sodium hypochlorite (<20 mg / l)
|
liquid, non-corrosive
to stainless steel
|
solution of sodium hypochlorite (<300 g / l)
|
liquid, non-corrosive
to titanium alloys
|
Deviations
|
5-6%
|
3%
|
10%
|
3%
|
Data communcations
|
n/a
|
Profibus
|
n/a
|
Modbus TCP
|
Power
|
220 VAC
|
24 VDC
|
220 VAC
|
220 VAC or
battery 24 VDC
|
Estimated price, units
|
1 530
|
400
|
450
|
150 - 300
|
4. Applications
- water treatment, disinfection of drinking water
- sugar and alcohol (the density measurement of sugar syrup, the alcohol content in alcohol)
- food production (measured density of the condensed milk, lactose, yoghurt, dairy acids, oils, extracts, starch solutions, glucose syrup, tomato paste, juices)
- brewing (beer imerenie density)
- chemicals (measurement of density and concentration of acid solution, caustic alkali solutions, detergent solutions)
- petrochemical industry (measurement of the density of oils, extracts, diesel fuel, petroleum products, gasoline, kerosene)
1 device production time: 1-2 weeks